High-performance liquid chromatographic analysis of commonly used drugs

Show full item record

Title: High-performance liquid chromatographic analysis of commonly used drugs
Author: Hider, Gregg C.
Abstract: This research project had two major goals. Selected members of six groups of clinically useful drugs were obtained to develop and optimize reverse-phase isocratic high-performance liquid chromatographic (HPLO separations using the Hewlett Packard Series 1090 M HPLC. All optimized methods displayed excellent resolution between compounds within a method. Each optimized drug assay developed, displayed good linearity and precision. The second and primary goal of this research project was to provide RIT with a teaching tool. These optimized separations were incorporated into six laboratory experiments, written in a format that RIT's undergraduate students of Medical Technology could easily follow and complete in one three hour student laboratory session. Each laboratory experiment is characterized by being a fast, simple, easy, and reliable reverse phase isocratic HPLC separation of select members of these drug groups that can be identified and quantitated in serum or plasma.
Record URI: http://hdl.handle.net/1850/10693
Date: 1988-05

Files in this item

Files Size Format View
GHiderThesis05-1988.pdf 3.095Mb PDF View/Open

The following license files are associated with this item: