Frequency dependent effective cell volume estimation for circular disk transducers using Lommel diffraction formulation

Show full item record

Title: Frequency dependent effective cell volume estimation for circular disk transducers using Lommel diffraction formulation
Author: Rao, Navalgund; Lai, Di
Abstract: In medical ultrasonic imaging the signal reflected from the tissue often has a random character to it. It is believed that the random nature of the tissue scattering microstructure is responsible for the stochastic nature of the echo signal. Chen, et. a!. have proposed a signal processing scheme that is based on the statistical moments calculated on the Fourier transform of the time gated echo signal. The theory requires the knowledge of a frequency-dependent effective cell volume term. This paper describes the use of a closed form expression (Lommel diffraction formulation) for this purpose. Our simulation results suggest that reliable estimation of the cell volume is possible only when the time duration of the excitation pulse is small compared to the time gate length.
Description: Copyright 2001 SPIE Medical Imaging proceedings.
Record URI: http://hdl.handle.net/1850/7843
Date: 2001

Files in this item

Files Size Format View
NRaoProceeding2001.pdf 458.4Kb PDF View/Open

The following license files are associated with this item: