Evaluation of the instrumentation laboratory microcentrifugal analyzer III

Show full item record

Title: Evaluation of the instrumentation laboratory microcentrifugal analyzer III
Author: McCall, Karin B.
Abstract: The Instrumentation Laboratory's MicroCentrifugal Analyzer (MCA) was evaluated for precision, linearity and accuracy for albumin, alanine transaminase (ALT), aspartate transaminase (AST) , alkaline phosphatase, creatine kinase (CK), cerebrospinal fluid (CSF) total protein, cholesterol, glucose, gamma-glutamyl transpeptidase (GGT), lactic dehydrogenase (LDH) , phosphorous, total protein, and triglyceride. Within-run, within-day, day-to-day, and total precision was comparable to or exceeded manufacturer's claims and was comparable to the precision in existing in-house tests. A comparison of methods for the MicroCentrifugal Analyzer with those used on the Technicon SMA 12/60, the Abbott ABA-100 and the Kodak Ektachem 400 showed very good correlation with minimal random error - As a result of this study, the MicroCentrifugal Analyzer was implemented in this laboratory as a replacement instrument for the ABA 100 and as a backup instrument for the SMA 12/60 and Ektachem 400. A new activated creatine kinase kit was adopted and the existing cerebrospinal fluid total protein biuret method was replaced with an automated Coomassie blue method.
Record URI: http://hdl.handle.net/1850/9635
Date: 1988-10

Files in this item

Files Size Format View
KMcCallThesis10-1988.pdf 1.725Mb PDF View/Open

The following license files are associated with this item: